Published October 2008 | Version v1
Journal article

Multipoint Thomson scattering diagnostic for the Madison Symmetric Torus reversed-field pinch

  • 1. Department of Physics, University of Wisconsin-Madison, 1150 University Avenue, Madison, Wisconsin 53706 (United States)

Description

The multipoint Thomson scattering diagnostic on the Madison Symmetric Torus (MST) is now fully operational with 21 spatial points, which cover the entire minor radius. Four full electron temperature profiles can be obtained during each MST discharge, with a variable delay between each profile. This system overcomes challenges that arise from the unique machine design, location, and plasma characteristics of MST. The machine design limits the maximum porthole diameter to 11.4 cm, requiring a compact, re-entrant, seven element lens for scattered light collection. Limited space near MST necessitates a long beam path for the two Nd:YAG lasers requiring a remote beam line adjustment system to suppress drift in the beam position due to thermal expansion of the building. Due to the remote location of the laser head, substantial design effort was put into the creation of a set of safety interlocks for the laser system. The dynamic nature of MST plasmas and the wide range of operating space require a versatile scattered light detection system consisting of filter polychromators with temperature controlled avalanche photodiode detectors. We also implement an insertable integrating sphere, which travels along the laser beam path through the vacuum vessel, for the alignment of both the fiber optics and the lasers.

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
79
Journal Issue
10
Journal Page Range
p. 10E733-10E733.5
ISSN
0034-6748
CODEN
RSINAK

Conference

Title
17. topical conference on high-temperature plasma diagnostics
Acronym
HTPD08
Dates
11-15 May 2008
Place
Albuquerque, NM (United States)

Optional Information

Notes
(c) 2008 American Institute of Physics