A study of Rhizophora spp wood phantom for dosimetric purposes using high-energy photon and electron beams
Creators
Description
Previous scattering and depth-dose investigations involving use of the Malaysian hardwood Rhizophora spp have shown this medium to produce good agreement with measurements made in water. Present study extends the comparison, now including measurements of percentage depth-dose made for photons at 6 MV and 5 and 12 MeV electron beams. For the 6 MV photon and 5 MeV electron beams, discrepancies between percentage depth-dose for Rhizophora spp and water, at all depths, are found to be within 2.6 and 2.4% respectively. At 12 MeV electron energies, measured percentage depth-doses in Rhizophora spp beyond 3.5 cm depth are found to be in significant discord with those for water. The absorbed dose in water measured in Rhizophora spp at dmax for all three beams produces discrepancies of no more than 1.1% when compared with measurements made in water
Additional details
Identifiers
- PII
- S0969804301000574;
Publishing Information
- Journal Title
- Applied Radiation and Isotopes
- Journal Volume
- 55
- Journal Issue
- 3
- Journal Page Range
- p. 297-302
- ISSN
- 0969-8043
- CODEN
- ARISEF
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- Canada
- INIS RN
- 33030285
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- DEPTH DOSE DISTRIBUTIONS; DOSIMETRY; ELECTRON BEAMS; PHANTOMS; PHOTON BEAMS; RHIZOPUS
- Descriptors DEC
- BEAMS; EUMYCOTA; FUNGI; LEPTON BEAMS; MOCKUP; PARTICLE BEAMS; PLANTS; RADIATION DOSE DISTRIBUTIONS; SPATIAL DOSE DISTRIBUTIONS; STRUCTURAL MODELS
Optional Information
- Copyright
- Copyright (c) 2001 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.