Analytic method for geometrical parameter correction of planar HPGe detector

  • 1. Northwest Institute of Nuclear Technology, Xi'an (China)

Description

A numerical integration formula was introduced to calculate the response of planar HPGe detector to photons emitted from point source. Then the formula was used to correct the geometrical parameter of planar HPGe detector. 241Am and 137Cs point sources were placed at a certain distance (1-20 cm) away from entrance window to get the corresponding detection efficiency. The detection parameters were calculated in weighted least square fitting using the formula with the experimental efficiencies as formula results. This correction method was accurate and timesaving. The simulation result from MCNP using the corrected parameters shows that the relative deviations between simulation and experimental efficiencies are less than 1% for 59.5 and 661.6 keV photons with the distance of 1-20 cm. (authors)

Additional details

Publishing Information

Journal Title
Atomic Energy Science and Technology
Journal Volume
47
Journal Issue
10
Journal Page Range
p. 1883-1887
ISSN
1000-6931

Optional Information

Notes
5 figs., 3 tabs., 7 refs.