An anthropomorphic phantom study of visualisation of surgical clips for partial breast irradiation (PBI) setup verification
Creators
- 1. Department of Radiation Oncology, Vancouver Centre, Vancouver, BC (Canada)
- 2. Department of Surveillance and Outcomes Unit, Vancouver Centre, Vancouver, BC (Canada)
- 3. Department of Radiation Therapy, Vancouver Centre, Vancouver, BC (Canada)
- 4. Department of Medical Physics, Vancouver Centre, Vancouver, BC (Canada)
Description
Surgical clips were investigated for partial breast image-guided radiotherapy (IGRT). Small titanium clips were insufficiently well visualised. Medium tantalum clips were best for megavoltage IGRT and small tantalum clips were best for floor mounted kilovoltage IGRT (ExacTracTM). Both small tantalum and medium titanium clips were suitable for isocentric kilovoltage IGRT
Availability note (English)
Available from http://dx.doi.org/10.1016/j.radonc.2008.03.011Additional details
Identifiers
- DOI
- 10.1016/j.radonc.2008.03.011;
- PII
- S0167-8140(08)00158-8;
Publishing Information
- Journal Title
- Radiotherapy and Oncology
- Journal Volume
- 90
- Journal Issue
- 1
- Journal Page Range
- p. 56-59
- ISSN
- 0167-8140
- CODEN
- RAONDT
INIS
- Country of Publication
- Ireland
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 40046120
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- IMAGES; IRRADIATION; MAMMARY GLANDS; PHANTOMS; RADIOTHERAPY; SURGERY; TANTALUM; TITANIUM; VERIFICATION
- Descriptors DEC
- BODY; ELEMENTS; GLANDS; MEDICINE; METALS; MOCKUP; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; REFRACTORY METALS; STRUCTURAL MODELS; THERAPY; TRANSITION ELEMENTS
Optional Information
- Copyright
- Copyright (c) 2008 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.