Published 1984 | Version v1
Report Open

New method of proportional counter feedback biasing for wide-range radiation dose-rate monitors

Description

A prototypic wide-range radiation dose-rate monitor for civil defense applications has been developed and tested. The specified dose-rate range (0 to 500 R/h) was displayed on a single readout scale by using feedback-controlled biasing of a proportional counter. This new method is based on controlling the avalanche multiplication factor (gas gain) of the counter by varying its bias voltage in response to its measured output current (i.e., detected dose rate). The counter output current varies between 0 and 1.5 nA in a quasi-logarithmic response to dose rates between 0 and 500 R/h. The corresponding values of gas gain and bias voltage range from 1 to 300 and 200 to 1900 V respectively

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS, PC A02/MF A01 as DE85002595.

Files

16046558.pdf

Files (99.0 kB)

Name Size Download all
md5:e5b5e60e3a939cb29524407c0fad1eba
99.0 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
4 p.
Report number
CONF-841007--19

Conference

Title
Nuclear science symposium.
Dates
31 Oct - 2 Nov 1984.
Place
Orlando, FL (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
16046558
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
AMPLIFIERS; DOSE RATEMETERS; FEEDBACK; GAIN; PROPORTIONAL COUNTERS
Descriptors DEC
EQUIPMENT; MEASURING INSTRUMENTS; RADIATION DETECTORS