Published 1984
| Version v1
Report
Open
New method of proportional counter feedback biasing for wide-range radiation dose-rate monitors
Description
A prototypic wide-range radiation dose-rate monitor for civil defense applications has been developed and tested. The specified dose-rate range (0 to 500 R/h) was displayed on a single readout scale by using feedback-controlled biasing of a proportional counter. This new method is based on controlling the avalanche multiplication factor (gas gain) of the counter by varying its bias voltage in response to its measured output current (i.e., detected dose rate). The counter output current varies between 0 and 1.5 nA in a quasi-logarithmic response to dose rates between 0 and 500 R/h. The corresponding values of gas gain and bias voltage range from 1 to 300 and 200 to 1900 V respectively
Availability note (English)
MF available from INIS under the Report Number; Available from NTIS, PC A02/MF A01 as DE85002595.Files
16046558.pdf
Files
(99.0 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:e5b5e60e3a939cb29524407c0fad1eba
|
99.0 kB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 4 p.
- Report number
- CONF-841007--19
Conference
- Title
- Nuclear science symposium.
- Dates
- 31 Oct - 2 Nov 1984.
- Place
- Orlando, FL (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 16046558
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- AMPLIFIERS; DOSE RATEMETERS; FEEDBACK; GAIN; PROPORTIONAL COUNTERS
- Descriptors DEC
- EQUIPMENT; MEASURING INSTRUMENTS; RADIATION DETECTORS