Published August 1991 | Version v1
Journal article

An anterior appositional electron field technique with a hanging lens block in orbital radiotherapy: A dosimetric study

  • 1. Michael Reese/University of Chicago, IL (USA)

Description

A technique for orbital radiotherapy is presented consisting of an anterior, appositional electron beam with a hanging lens block. The beam was modified by introducing two 1.6 mm thick plastic spoilers, at about 3 cm and 15 cm from the lens, to boost in-scattering of electrons under the block. The 9 mm diameter, 2 cm long stainless steel cylindrical block was suspended 0.5-1.0 cm above the eye. The authors performed film, TLD (Thermo Luminescent Dosimetry), and diode dosimetry to determine the dose fill-in behind the lens. The introduction of the spoilers dramatically changed the dose distribution. The maximum dose under the block increased from 66% to 85% of the open field dose. Moreover, the dose to the posterior surface of the globe directly underneath the block, at a depth of 3 cm, increased from 48% to 76% of maximum dose, while the dose to the lens was still below 20%. This is a simple and easily reproducible treatment and is an improvement on a previously described technique. The dose distribution is adequate for cases where the target volume surrounds and is posterior to the globe

Additional details

Publishing Information

Journal Title
International Journal of Radiation Oncology, Biology and Physics
Journal Volume
21
Journal Issue
3
Series
Int. J. Radiat. Oncol., Biol. Phys.
Journal Page Range
795-804
ISSN
0360-3016
CODEN
IOBPD

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
23005336
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
CRYSTALLINE LENS; ELECTRON BEAMS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIATION PROTECTION; RADIOTHERAPY; THERMOLUMINESCENT DOSIMETRY
Descriptors DEC
BEAMS; BODY; DOSIMETRY; EYES; LEPTON BEAMS; MEDICINE; ORGANS; PARTICLE BEAMS; SENSE ORGANS; THERAPY