Published November 2005 | Version v1
Journal article

Properties of novel MSGC of diamond film for x-ray

  • 1. Shanghai Univ., Shanghai (China). School of Materials Science and Engineering

Description

MSGC was prepared on diamond film/silicon composite substrate using Cr/Au as electrodes materials. Micro-strip electrodes were photo-etched by modern microelectronic technology. Energy resolution and pulse signal of 5.9 keV 55Fe X-ray using Ar + CH4 mixture as working gas under room temperature 101.0 kPa with different voltage were measured by a Multi-channel energy spectrometer. Results indicated that high signal-noise ratio, counting rate (103 Hz of the MSGC were obtained. When Ar:CH4=90:10, drift voltage, cathode voltage and anode voltage was -1000 V, -650 V and 0 V, respectively, the energy resolution reached 12.3%. (authors)

Additional details

Additional titles

Augmented title (English)
The MSGC is stated for Micro-Strip Gas Chember

Publishing Information

Journal Title
Nuclear Electronics and Detection Technology
Journal Volume
25
Journal Issue
6
Journal Page Range
p. 750-753
ISSN
0258-0934

Optional Information

Notes
6 figs., 14 refs.