Validating dose rate calibration of radiotherapy photon beams through IAEA/WHO postal audit dosimetry service
Creators
- 1. The Aga Khan University Hospital, Karachi (Pakistan). Dept. of Radiation Oncology
Description
In external beam radiation therapy (EBRT), the quality assurance (QA) of the radiation beam is crucial to the accurate delivery of the prescribed dose to the patient. One of the dosimetric parameters that require monitoring is the beam output, specified as the dose rate on the central axis under reference conditions. The aim of this project was to validate dose rate calibration of megavoltage photon beams using the International Atomic Energy Agency (IAEA)/World Health Organisation (WHO) postal audit dosimetry service. Three photon beams were audited: a 6 MV beam from the low-energy linac and 6 and 18 MV beams from a dual high-energy linac. The agreement between our stated doses and the IAEA results was within 1% for the two 6 MV beams and within 2% for the 18 MV beam. The IAEA/WHO postal audit dosimetry service provides an independent verification of dose rate calibration protocol by an international facility. (author)
Additional details
Publishing Information
- Journal Title
- Journal of the Pakistan Medical Association
- Journal Volume
- 62
- Journal Issue
- 5
- Journal Page Range
- p. 490-492
- ISSN
- 0030-9982
INIS
- Country of Publication
- Pakistan
- Country of Input or Organization
- Pakistan
- INIS RN
- 43070485
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- AUDITS; CALIBRATION; DOSE RATES; DOSIMETRY; IAEA; PHOTON BEAMS; QUALITY ASSURANCE; RADIOTHERAPY; WHO
- Descriptors DEC
- BEAMS; INTERNATIONAL ORGANIZATIONS; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY