High precision straw tube chamber with cathode readout
Description
The high precision straw chamber with cathode readout was constructed and investigated. The 10 mm straws were made of aluminized mylar strip with transparent longitudinal window. The X coordinate information has been taken from the cathode strips as induced charges and investigated via centroid method. The spatial resolution σ=120 μm has been obtained with signal/noise ratio about 60. The possible ways for improving the signal/noise ratio have been described. 7 refs.; 8 figs
Availability note (English)
MF available from INIS under the Report Number.Files
25001025.pdf
Files
(164.4 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:02f09c6c44e3fec4b5aaa281afc72f41
|
164.4 kB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 8 p.
- Report number
- JINR-E--1-92-199
INIS
- Country of Publication
- Russian Federation
- Country of Input or Organization
- Russian Federation
- INIS RN
- 25001025
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACCURACY; CHARGE DISTRIBUTION; COUNTING RATES; KEV RANGE 01-10; MULTIWIRE PROPORTIONAL CHAMBERS; MYLAR; POSITION SENSITIVE DETECTORS; READOUT SYSTEMS; SIGNAL-TO-NOISE RATIO; SPATIAL DISTRIBUTION; SPATIAL RESOLUTION; X-RAY DETECTION
- Descriptors DEC
- DETECTION; DISTRIBUTION; ENERGY RANGE; ESTERS; KEV RANGE; MEASURING INSTRUMENTS; ORGANIC COMPOUNDS; POLYESTERS; PROPORTIONAL COUNTERS; RADIATION DETECTION; RADIATION DETECTORS; RESOLUTION
Optional Information
- Notes
- Submitted to Nucl. Instrum. Methods.