Published January 1996 | Version v1
Report Open

TSEE characteristics of LiF:M,C,P detectors

Creators

  • 1. Institute of Isotopes of the Hungarian Academy of Sciences, Budapest (Hungary)

Description

LiF based thermoluminescence detectors are most attractive amongst various thermoluminescent dosimetric materials because of their low energy dependence and near tissue equivalence. The ultra sensitive LiF:Mg,Cu,P detectors now are available on commercial scale and widely used for personal dosimetric and environmental monitoring purposes. The Chinese type GR-200 and MCPN LiF:Mg,Cu,P IPN, Krakow TL detectors are similar in their main characteristics. The effective atomic number is 8.2, the emission spectrum maxima is 4000 A, typical sensitivity at 60Co relative to TLD-100 is 15-50, the overresponse at 30 keV/60Co is 1.06 and the fading is negligible, less 5%/year. The useful dose range of this material is 1μ y-10 Gy. The high sensitivity of this material allow to prepare thin sensitive layers by which nearly energy independent beta dose measurement can be performed. LiF:MCP detectors are very delicate regarding their heat treatment. Typically 10 mins at 240 deg. C is the annealing suggested even though annealing up to 270 deg. C may give good reproducibility, too. Up to 12-15 % loss in sensitivity was observed on repeated measurements up to 50 cycles using 10 mins annealing at 240 deg. C. Annealing up to 400 deg. C for 2 hrs resulted the shift of the main dosimetric peak together with a significant decrease of sensitivity. LiF based dosimetric materials besides TL exhibit thermally stimulated exoelectron emission, TSEE, phenomenon, too. The analysis of TL and TSEE glow curve structure may reveal interrelation between peaks therefore the physical process responsible for emissions. The main purpose of present work was to measure and investigate both TL and TSEE responses of LiF:Mg,Cu,P detectors using different heat treatment

Files

37004368.pdf

Files (207.9 kB)

Name Size Download all
md5:d55f7dd9126c220c31a5a69954252fa8
207.9 kB Preview Download
Part of:
Symposium on radiation protection in neighbouring countries in Central Europe - 1995. Proceedings

Additional details

Publishing Information

Imprint Title
Symposium on radiation protection in neighbouring countries in Central Europe - 1995. Proceedings
Imprint Pagination
478 p.
Journal Page Range
p. 285-288
Report number
INIS-XA-C--071

Conference

Title
1995 IRPA regional congress on radiation protection in Central Europe
Dates
4-8 Sep 1995
Place
Portoroz (Slovenia)

INIS

Country of Publication
International Atomic Energy Agency (IAEA)
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
37004368
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
LITHIUM FLUORIDES; PERSONNEL DOSIMETRY; RADIATION MONITORING; RADIATION MONITORS; THERMOLUMINESCENT DOSEMETERS; THERMOLUMINESCENT DOSIMETRY
Descriptors DEC
ALKALI METAL COMPOUNDS; DOSEMETERS; DOSIMETRY; FLUORIDES; FLUORINE COMPOUNDS; HALIDES; HALOGEN COMPOUNDS; LITHIUM COMPOUNDS; LITHIUM HALIDES; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; MONITORING; MONITORS

Optional Information

Contract/Grant/Project number
Contract ERBFI13CTP 920032
Notes
3 refs, 5 figs Imprint:Pages missing in original document