Published 1977
| Version v1
Report
parameters influencing electronic noise in a typical microdosimetric arrangement
Creators
Description
Causes of electronic noise in a microdosimetric arrangement consisting of a proportional counter, counter base and connectors, the preamplifier, and the amplifier are described. (1) Leads, spring loaded contacts and pins, and the wiring on the preamplifier board, (2) the insulating material, (3) the FET transistor, (4) RC shaping of pulses, and (5) cooling of FET and feedback resistor are listed
Additional details
Publishing Information
- Imprint Title
- Annual report on research project, October 1, 1976--September 30, 1977
- Journal Page Range
- p. 83-87.
- Report number
- COO--3243-6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 10438233
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ELECTRICAL INSULATORS; FIELD EFFECT TRANSISTORS; MICRODOSIMETRY; NOISE; PROPORTIONAL COUNTERS
- Descriptors DEC
- DOSIMETRY; ELECTRICAL EQUIPMENT; MEASURING INSTRUMENTS; RADIATION DETECTORS; SEMICONDUCTOR DEVICES; TRANSISTORS