Published December 2011 | Version v1
Journal article

Time-resolved soft x-ray absorption setup using multi-bunch operation modes at synchrotrons

  • 1. Sincrotrone Trieste, S.S. 14 km 163.5, Area Science Park, 34149 Basovizza (Italy)
  • 2. Department of Physics, University of Trieste, via A. Valerio 2, 34127, Trieste (Italy)
  • 3. IOM-CNR, TASC laboratory, S.S. 14 km 163.5, Area Science Park, 34149 Basovizza (Italy)

Description

Here, we report on a novel experimental apparatus for performing time-resolved soft x-ray absorption spectroscopy in the sub-ns time scale using non-hybrid multi-bunch mode synchrotron radiation. The present setup is based on a variable repetition rate Ti:sapphire laser (pump pulse) synchronized with the ∼500 MHz x-ray synchrotron radiation bunches and on a detection system that discriminates and singles out the significant x-ray photon pulses by means of a custom made photon counting unit. The whole setup has been validated by measuring the time evolution of the L3 absorption edge during the melting and the solidification of a Ge single crystal irradiated by an intense ultrafast laser pulse. These results pave the way for performing synchrotron time-resolved experiments in the sub-ns time domain with variable repetition rate exploiting the full flux of the synchrotron radiation.

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
82
Journal Issue
12
Journal Page Range
p. 123109-123109.6
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2011 American Institute of Physics