Published May 1985 | Version v1
Journal article

Stabilization of scintillation detector by reference signal from radioactive source

  • 1. Sci-Res. Inst. of Electron Introscopy, Tomsk

Description

This paper examines the possibility of stabilization of a scintillation detector using an additional detector whose scintillator with respect to the former performs the function of a target, which generates secondary radiation. It is experimentally established that a regulation time constant of 10 sec ensures constancy of the transfer characteristic of the detector with an error of about 1.5% for disturbing effects that correspond to 20% of this characteristic

Additional details

Publishing Information

Journal Title
Instrum. Exp. Tech. (Engl. Transl.)
Journal Volume
27
Journal Issue
6
Series
Instrum. Exp. Tech. (Engl. Transl.).
Journal Page Range
1363-1365
ISSN
0020-4412
CODEN
INETA

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
18006194
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
CALIBRATION; ERRORS; RADIATION SOURCES; SCINTILLATION COUNTERS; STABILITY; TARGETS; TIMING PROPERTIES
Descriptors DEC
MEASURING INSTRUMENTS; RADIATION DETECTORS

Optional Information

Notes
Translated from Pribory i Tekhnika Eksperimenta, No. 6, pp. 58-61, November-December, 1984.. Cover-to-cover translation of Pribory i Tekhnika Ehksperimenta (USSR).