Description
A passive neutron dosemeter was designed to be used in mixed radiation fields. The design was carried out using Monte Carlo method. The dosemeter model was a 25.4 cm-diameter polyethylene sphere with a thermoluminescent dosemeter, TLD600, located at the sphere center. This model was irradiated with 50 monoenergetic neutron sources with energies from 10-8 to 20 MeV. A 506.71 cm2-area disk was used to model the source term whose center was located at 100 cm from polyethylene sphere's center. The dosemeter response was compared with the responses of SNOOPY, Harwell 95/0075 and PNR-4. With these responses it was calculated the dosemeter responses for 252Cf, 252Cf/D2O and 239PuBe neutron sources. The passive dosemeter relative response has the same shape of SNOOPY, Harwell 95/0075 and PNR-4 dosemeters. Due to the type of thermal neutron detector used in the passive dosemeter the absolute response per unit fluence, is lower than the absolute response of SNOOPY, Harwell 95/0075 and PNR-4 dosemeters. However the passive dosemeter response in function of the average neutron energy of the 252Cf, 252Cf/D2O and 239PuBe neutron energy was more linear
Additional details
Identifiers
- DOI
- 10.1063/1.1512056;
Publishing Information
- Journal Title
- AIP Conference Proceedings
- Journal Volume
- 630
- Journal Issue
- 1
- Journal Page Range
- p. 215-219
- ISSN
- 0094-243X
- CODEN
- APCPCS
Conference
- Title
- 6. Mexican symposium on medical physics
- Dates
- 20-22 Mar 2002
- Place
- Mexico City (Mexico)
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 34067670
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference, Numerical Data
- Descriptors DEI
- DESIGN; DOSEMETERS; MONTE CARLO METHOD; NEUTRON BEAMS; NEUTRON DETECTORS; NEUTRON DOSIMETRY; PROTON DOSIMETRY; THEORETICAL DATA
- Descriptors DEC
- BEAMS; CALCULATION METHODS; DATA; DOSIMETRY; INFORMATION; MEASURING INSTRUMENTS; NUCLEON BEAMS; NUMERICAL DATA; PARTICLE BEAMS; RADIATION DETECTORS
Optional Information
- Notes
- (c) 2002 American Institute of Physics