Published October 9, 2002 | Version v1
Journal article

Passive neutron dosemeter design

  • 1. Unidades Academicas de Ingenieria, Apdo. Postal 336, 98000 Zacatecas, Zac (Mexico)
  • 2. Unidades Academicas de Estudios Nucleares, Apdo. Postal 336, 98000 Zacatecas, Zac (Mexico)
  • 3. Unidades Academicas de Matematicas de la UAZ, Apdo. Postal 336, 98000 Zacatecas, Zac (Mexico)

Description

A passive neutron dosemeter was designed to be used in mixed radiation fields. The design was carried out using Monte Carlo method. The dosemeter model was a 25.4 cm-diameter polyethylene sphere with a thermoluminescent dosemeter, TLD600, located at the sphere center. This model was irradiated with 50 monoenergetic neutron sources with energies from 10-8 to 20 MeV. A 506.71 cm2-area disk was used to model the source term whose center was located at 100 cm from polyethylene sphere's center. The dosemeter response was compared with the responses of SNOOPY, Harwell 95/0075 and PNR-4. With these responses it was calculated the dosemeter responses for 252Cf, 252Cf/D2O and 239PuBe neutron sources. The passive dosemeter relative response has the same shape of SNOOPY, Harwell 95/0075 and PNR-4 dosemeters. Due to the type of thermal neutron detector used in the passive dosemeter the absolute response per unit fluence, is lower than the absolute response of SNOOPY, Harwell 95/0075 and PNR-4 dosemeters. However the passive dosemeter response in function of the average neutron energy of the 252Cf, 252Cf/D2O and 239PuBe neutron energy was more linear

Additional details

Identifiers

Publishing Information

Journal Title
AIP Conference Proceedings
Journal Volume
630
Journal Issue
1
Journal Page Range
p. 215-219
ISSN
0094-243X
CODEN
APCPCS

Conference

Title
6. Mexican symposium on medical physics
Dates
20-22 Mar 2002
Place
Mexico City (Mexico)

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
34067670
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference, Numerical Data
Descriptors DEI
DESIGN; DOSEMETERS; MONTE CARLO METHOD; NEUTRON BEAMS; NEUTRON DETECTORS; NEUTRON DOSIMETRY; PROTON DOSIMETRY; THEORETICAL DATA
Descriptors DEC
BEAMS; CALCULATION METHODS; DATA; DOSIMETRY; INFORMATION; MEASURING INSTRUMENTS; NUCLEON BEAMS; NUMERICAL DATA; PARTICLE BEAMS; RADIATION DETECTORS

Optional Information

Notes
(c) 2002 American Institute of Physics