Published January 1982
| Version v1
Journal article
A technique of scatter and glare correction for videodensitometric studies in digital subtraction videoangiography
Creators
- 1. Department of Medical Physics, University of Wisconsin, Madison
Description
The logarithmic amplification of video signals and the availability of data in digital form make digital subtraction videoangiography a suitable tool for videodensitometric estimation of physiological quantities. A system for this purpose was implemented with a digital video image processor. However, it was found that the radiation scattering and veiling glare present in the image-intensified video must be removed to make meaningful quantitations. An algorithm to make such a correction was developed and is presented. With this correction, the videodensitometry system was calibrated with phantoms and used to measure the left ventricular ejection fraction of a canine heart
Additional details
Publishing Information
- Journal Title
- Radiology
- Journal Volume
- 142
- Journal Issue
- 1
- Series
- Radiology.
- Journal Page Range
- 209-213
- ISSN
- 0033-8419
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 13703214
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ALGORITHMS; BIOMEDICAL RADIOGRAPHY; BLOOD VESSELS; CALIBRATION; DENSIMETERS; DOGS; HEART; IMAGES; OPTIMIZATION; PHANTOMS; SCATTERING
- Descriptors DEC
- ANIMALS; BODY; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; MAMMALS; MEASURING INSTRUMENTS; MEDICINE; MOCKUP; ORGANS; STRUCTURAL MODELS; VERTEBRATES