Published 2017
| Version v1
Journal article
Filtration influence in a constant potential X-ray machine peak voltage measurements
- 1. Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)
- 2. Instituto Federal de Educacao, Ciencia e Tecnologia da Bahia (IFBA), Salvador (Brazil)
Description
This work shows the peak voltage measurements for several beam filtrations used in diagnostic radiology, using two types of non-invasive detectors; a voltage meter and a high-resolution spectrometer. The technique chosen for the voltage peak measurements with the spectrometer was the endpoint. The results were compared to the measured ones and showed good similarity to the nominal values. However the voltage meter detector used in this work presented errors for heavier filtrations. (author)
Availability note (English)
Available from http://www.bjrs.org.br/revista/index.php/REVISTA/article/view/261/214Additional details
Publishing Information
- Journal Title
- Brazilian Journal of Radiation Sciences
- Journal Volume
- 5
- Journal Issue
- 2
- Journal Page Range
- p. 1-9
- ISSN
- 2319-0612
INIS
- Country of Publication
- Brazil
- Country of Input or Organization
- Brazil
- INIS RN
- 48077176
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- BEAMS; CALIBRATION; ELECTRIC POTENTIAL; PEAKS; RADIOLOGY; SEMICONDUCTOR DETECTORS; SPECTROSCOPY; X-RAY DOSIMETRY; X-RAY SOURCES
- Descriptors DEC
- DOSIMETRY; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; RADIATION DETECTORS; RADIATION SOURCES