Published January 1982 | Version v1
Journal article

Tissue-air ratio and its associated 'in-air' reference minimal phantom

Creators

  • 1. National Research Council of Canada, Ottawa, Ontario. Div. of Physics

Description

A problem is raised concerning a revised equation published in 1981 relating tissue-air ratio to per cent depth dose for an incident photon beam. The problem mainly involves the required size of the reference minimal phantom that is irradiated free-in-air to yield the reference absorbed dose as a tissue-air ratio. If the radius of the minimal phantom is significantly less than the range of the most energetic secondary electrons, the reference absorbed dose is indeterminate. It is shown that the small phantom size assumed in the 1981 publication could result, even if no electrons contaminate the incident photon beam, in a large discrepancy. (U.K.)

Additional details

Publishing Information

Journal Title
Phys. Med. Biol.
Journal Volume
27
Journal Issue
1
Series
Phys. Med. Biol.
Journal Page Range
153-154
ISSN
0031-9155

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
13677952
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
DEPTH DOSE DISTRIBUTIONS; DIMENSIONS; ERRORS; GAMMA DOSIMETRY; PHANTOMS
Descriptors DEC
DOSIMETRY; MOCKUP; RADIATION DOSE DISTRIBUTIONS; SPATIAL DOSE DISTRIBUTIONS; STRUCTURAL MODELS