The feasibility of in vivo detection of gadolinium by prompt gamma neutron activation analysis following gadolinium-based contrast-enhanced MRI

  • 1. Department of Medical Physics and Applied Radiation Sciences, McMaster University, Hamilton, Ontario, L8S-4K1 (Canada)

Description

The feasibility of using the McMaster University in vivo prompt gamma neutron activation analysis system for the detection of gadolinium has been investigated. Phantoms have been developed for the kidney, liver, and the leg muscle. The initial detection limits are determined to be 7.2±0.3 ppm for the kidney, 3.0±0.1 ppm for the liver, and 2.33±0.08 ppm for the lower leg muscle. A few system optimizations have been tested and show significant detection limit reduction from these initial values. The technique is promising and shows feasibility for in vivo studies of gadolinium retention.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2010.07.023

Additional details

Identifiers

DOI
10.1016/j.apradiso.2010.07.023;
PII
S0969-8043(10)00309-X;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
69
Journal Issue
1
Journal Page Range
p. 105-111