Estimation of radiation doses to patients and surgeons from various fluoroscopically guided orthopaedic surgeries
Creators
- 1. Medical Physics Dept., Agios Savvas Hospital, 171 Alexandras Avenue, 115 22 Athens (Greece)
- 2. Medical Physics Unit, 'Konstantopoulio - Agia Olga' Hospital, 3-5 Agias Olgas, Nea Ionia 142 33 Athens (Greece)
- 3. B Orthopaedic Dept., Univ. of Athens, 'Konstantopoulio - Agia Olga' Hospital, 3-5 Agias Olgas, Nea Ionia 142 33 Athens (Greece)
- 4. Radiology Dept., 'Konstantopoulio - Agia Olga' Hospital, 3-5 Agias Olgas, Nea Ionia 142 33 Athens (Greece)
Description
In this study, a mathematical method was used to estimate the entrance surface dose (ESD) to the patient and the scattered dose (Ds) to the operating surgeon during various fluoroscopically guided surgical orthopaedic procedures. For 204 patients, the procedure type, the fluoroscopy time and the highest tube potential and current values observed during fluoroscopy were recorded. For the most often performed procedures (intramedullary nailing of peritrochanteric fractures, open reduction and internal fixation of malleolar fractures and intramedullary nailing of diaphyseal fractures of the femur), the respective mean fluoroscopy times were 3.2, 1.5 and 6.3 min while the estimated mean ESDs were 183, 21 and 331 mGy, respectively. The estimated Ds rates for the hands, chest, thyroid, eyes, gonads and legs of the operating surgeon were on average to 0.103, 0.023, 0.013, 0.012, 0.066 and 0.045 mGy min-1, respectively, and compare well with the literature. The mathematical estimation of doses cannot replace actual measurements; however, it can be used for a preliminary assessment of the radiation dose levels during various surgical procedures, so that the operator, the surgeon and the rest of the medical staff involved could be aware of the associated radiation risk and the radiation protection measures required. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncm234Additional details
Identifiers
- DOI
- 10.1093/rpd/ncm234;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 128
- Journal Issue
- 1
- Journal Page Range
- p. 112-119
- ISSN
- 0144-8420
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 43120503
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- CHEST; EYES; FEMUR; FLUOROSCOPY; FRACTURES; HANDS; HUMAN POPULATIONS; LEGS; OCCUPATIONAL EXPOSURE; PATIENTS; RADIATION DOSES; RADIATION MONITORING; RADIATION PROTECTION; SURFACES; SURGERY; THYROID
- Descriptors DEC
- ARMS; BIOMEDICAL RADIOGRAPHY; BODY; DIAGNOSTIC TECHNIQUES; DOSES; ENDOCRINE GLANDS; FACE; FAILURES; GLANDS; HEAD; LIMBS; MEDICINE; MONITORING; NUCLEAR MEDICINE; ORGANS; POPULATIONS; RADIOLOGY; SENSE ORGANS; SKELETON
Optional Information
- Notes
- 20 refs