Influence of the profile of Iridium-192 wire in measurements of quality control for use in brachytherapy
Creators
- 1. Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)
Description
Brachytherapy is a method used in the treatment of cancerous tumors, by ionizing radiation produced by sources, introduced into the tumor area, this method seeks a more direct to the tumor, thereby maximizing the radiation dose to diseased tissue, while minimizing the dose to healthy tissues. One of the radionuclides used in brachytherapy is iridium-192. The Radiation Technology Center (CTR), of the Nuclear and Energy Research Institute (IPEN), has produced commercially, since 1998, iridium-192 wires used in low dose rate (LDR) brachytherapy. To produce this radionuclides, firstly an iridium-platinum wire is irradiated in the nuclear reactor IEA-R1 for 30 hours, with a neutron flux of 5x 1013 n cm-2 s-1, the wire is left to decay by 30 days to remove the main contaminants, and then goes through a quality control, before being sent to the hospital. In this quality control is checked the radiation homogeneity along each centimeter of the wire. To implement this procedure, is used a device, consisting of an ionization chamber, surrounded by a lead shield, with a samol 1 cm wide slit, linked to the ionization chamber is a voltage source, and a Keithley 617 electrometer, 2 minutes is the range used to measure the charge by the electrometer. The iridium wire is considered in accordance, when there is no variation higher than 5%, between the average measures and the maximum and minimum values. However, due to design features of the measurement system, the wire may appear to the detector through the slit in larger sizes than the ideal, improperly influencing the final quality control. This paper calculates the difference in size of these variations in profile, and their influence on the final count, it compares the actual values obtained, and describes the improvements, made in quality control procedures, that provided more accurate measurement data, analyzes the results, and suggests changes in devices, aimed at further improving the quality control of iridium-192 weirs produced at IPEN, and used in hospitals in Brazil. (author)
Files
46015547.pdf
Files
(480.9 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:324a6e40f7b8f8e41f8e144f2cd0da5c
|
480.9 kB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 6 p.
- Report number
- INIS-BR--14801
Conference
- Title
- international nuclear atlantic conference
- Acronym
- INAC 2013
- Dates
- 24-29 Nov 2013
- Place
- Recife, PE (Brazil)
INIS
- Country of Publication
- Brazil
- Country of Input or Organization
- Brazil
- INIS RN
- 46015547
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BRACHYTHERAPY; BRAZILIAN CNEN; ELECTROMETERS; ERRORS; IONIZATION CHAMBERS; IRIDIUM 192; IRRADIATION; NEUTRON FLUX; QUALITY CONTROL; WIRES
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BRAZILIAN ORGANIZATIONS; CONTROL; DAYS LIVING RADIOISOTOPES; ELECTRIC MEASURING INSTRUMENTS; ELECTRICAL EQUIPMENT; ELECTRON CAPTURE RADIOISOTOPES; EQUIPMENT; HEAVY NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; IRIDIUM ISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MEASURING INSTRUMENTS; MEDICINE; MINUTES LIVING RADIOISOTOPES; NATIONAL ORGANIZATIONS; NUCLEAR MEDICINE; NUCLEI; ODD-ODD NUCLEI; RADIATION DETECTORS; RADIATION FLUX; RADIOISOTOPES; RADIOLOGY; RADIOTHERAPY; THERAPY; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- 5 refs., 5 figs.