Determination of the effective dose for CT examinations and influence of the setup parameters

  • 1. Radiologische Physik, Universitaetsspital Basel (Switzerland)

Description

Purpose: CT examinations are considered to be high dose applications in radiation diagnostics. Consequently, radiation exposure and its potential reduction are an important issue. The aim of this paper is to determine effective doses for CT examinations under clinical conditions. The effective doses were determined from organ dose measurements with an optimal amount of effort. Materials and methods: the measured point doses in the various organs in an anthropomorphic phantom were weighted with respect to morphology and location in the radiation field and were averaged to obtain the organ doses. These doses together with the individual radiation sensitivities were able to be used to calculate the effective dose. The following examinations were taken into consideration: thorax, abdomen and skull. Only one parameter was changed for each measurement (the high voltage of the X-ray tube, the pitch, the collimation, or the automatic adjustment of the tube current to correct the individual absorption in the phantom). Results: the effective doses for the thorax examinations were between 5 and 7 mSv, and for the abdomen between 8 and 15 mSv depending on the technique and parameters. In some of the organs the organ doses can be in excess of 25 mSv. Our results were compared to the corresponding results of two commercial computer programs for dose calculations. The calculated effective doses generally showed lower values. (orig.)

Additional details

Additional titles

Original title (German)
Die Bestimmung der effektiven Dosen bei CT-Untersuchungen und deren Beeinflussung durch Einstellparameter

Publishing Information

Journal Title
RoeFo - Fortschritte auf dem Gebiete der Roentgenstrahlen und der Bildgebenden Verfahren
Journal Volume
179
Journal Issue
1
Journal Page Range
p. 38-45
ISSN
1438-9029
CODEN
RFGNDO

INIS

Country of Publication
Germany
Country of Input or Organization
Germany
INIS RN
38039956
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
COMPUTERIZED TOMOGRAPHY; DOSIMETRY; EFFICIENCY; PHANTOMS; RADIATION DOSES
Descriptors DEC
DIAGNOSTIC TECHNIQUES; DOSES; MOCKUP; STRUCTURAL MODELS; TOMOGRAPHY