Published January 27, 2014 | Version v1
Journal article

Hydrogen peroxide detection with quartz-enhanced photoacoustic spectroscopy using a distributed-feedback quantum cascade laser

  • 1. Department of Electrical and Computer Engineering, Rice University, 6100 Main Street, Houston, Texas 77005 (United States)
  • 2. Department of Civil and Environmental Engineering, Rice University, 6100 Main Street, Houston, Texas 77005 (United States)
  • 3. Dipartimento Interateneo di Fisica, Università e Politecnico di Bari, Via Amendola 173, Bari 70126 (Italy)
  • 4. Corning Incorporated, Corning, New York 14831 (United States)

Description

A quartz-enhanced photoacoustic spectroscopy sensor system was developed for the sensitive detection of hydrogen peroxide (H2O2) using its absorption transitions in the v6 fundamental band at ∼7.73 μm. The recent availability of distributed-feedback quantum cascade lasers provides convenient access to a strong H2O2 absorption line located at 1295.55 cm−1. Sensor calibration was performed by means of a water bubbler that generated titrated average H2O2 vapor concentrations. A minimum detection limit of 12 parts per billion (ppb) corresponding to a normalized noise equivalent absorption coefficient of 4.6 × 10−9 cm−1W/Hz1/2 was achieved with an averaging time of 100 s

Additional details

Identifiers

Publishing Information

Journal Title
Applied Physics Letters
Journal Volume
104
Journal Issue
4
Journal Page Range
p. 041117-041117.5
ISSN
0003-6951
CODEN
APPLAB

Optional Information

Notes
(c) 2014 AIP Publishing LLC