Comparison of Dw measurements by alanine and synthetic diamond dosimeters in photon beams with 1 cm * 1 cm field size

  • 1. Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, ENEA-INMRI C R Casaccia, Via Anguillarese 301, 00123 Roma, (Italy)
  • 2. Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, (Germany)

Description

For photon beam sizes lower than 2 cm * 2 cm no absorbed-dose-to-water (Dw) primary standards are available for determining the possible influence of field size on the dosimeter response. In this work, Dw measurements performed in 1 cm * 1 cm photon beams at PTB using alanine and synthetic diamond dosimeters were compared in order to assess the reliability of the two dosimetric systems in such a small field size. The ratio of Dw values measured by diamond dosimeters to Dw values measured by alanine was 0.991 ± 1.5% (1 SD) and 1.007 ± 1.3% (1 SD) for 6 MV and 10 MV photon beams, respectively. This shows that both systems are capable of accurately measuring Dw in small fields. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1088/0026-1394/49/5/S211