A six-degree-of-freedom robotic motion system for quality assurance of real-time image-guided radiotherapy
Creators
- 1. ACRF Image X Institute, Central Clinical School, The University of Sydney, Sydney, NSW (Australia)
- 2. School of Aerospace, Mechanical, and Mechatronic Engineering, University of Sydney, Sydney, NSW (Australia)
- 3. Northern Sydney Cancer Centre, Royal North Shore Hospital, Sydney, NSW (Australia)
Description
In this study we develop and characterise a six degree-of-freedom (6 DoF) robotic motion system for quality assurance of real-time image-guided radiotherapy techniques. The system consists of a commercially available robotic arm, an acrylic phantom with embedded Calypso markers, a custom base plate to mount the robot to the treatment couch, and control software implementing the appropriate sequence of transformations to reproduce measured tumour motion traces. The robotic motion system was evaluated in terms of the set-up and motion trace repeatability, static localization accuracy and dynamic localization accuracy. Four prostate, two liver and three lung motion traces, representing a range of tumor motion trajectories recorded in real patient treatments, were executed using the robotic motion system and compared with motion measurements from the clinical Calypso motion tracking system. System set-up and motion trace repeatability was better than 0.5 deg and 0.3 mm for rotation and translation, respectively. The static localization accuracy of the robotic motion system in the LR, SI and AP directions was 0.09 mm, 0.08 mm and 0.02 mm for translations, respectively, and 0.2°, 0.06° and 0.06° for rotations, respectively. The dynamic localization accuracy of the robotic motion system was <0.2 mm and <0.6° for translations and rotations, respectively. Thus, we have demonstrated the ability to accurately mimic rigid-body tumor motion using a robotically controlled phantom to provide precise geometric QA for advanced radiotherapy delivery approaches. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1361-6560/ab1935Additional details
Identifiers
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 64
- Journal Issue
- 10
- Journal Page Range
- [12 p.]
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 52003810
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- DEGREES OF FREEDOM; NEOPLASMS; QUALITY ASSURANCE; RADIOTHERAPY; ROBOTS; ROTATION; TRANSLATORS
- Descriptors DEC
- COMPUTER CODES; DISEASES; EQUIPMENT; MANAGEMENT; MEDICINE; MOTION; NUCLEAR MEDICINE; QUALITY MANAGEMENT; RADIOLOGY; THERAPY