Equipment standards for interventional cardiology
Creators
- 1. Dept. of Medical Physics and Bioengineering, St. James's Hospital, Garden Hill House, Dublin 8 (Ireland)
Description
Interventional radiology has seen rapid growth in cardiology and represents an alternative to hazardous surgery. Recently there has been a substantial growth in the number of procedures being performed and interventional cardiology (IC) procedures are the most common interventional procedures in Europe. Advances in imaging technology have facilitated the development of increasingly complex radiological IC equipment. Currently, the technology is developing at a rate ahead of supporting research, equipment standards and a regulatory framework. International standards play a key role in the design, manufacture and performance of radiological IC equipment. A survey of 12 IC systems (15 imaging chains) was conducted in Irish hospitals. The aim of the study was to assess the imbalance between rapidly advancing technology and existing standards and to propose recommendations for new IC equipment standards. The results demonstrate the need for definitive equipment requirements and standardisation in the design, manufacture, acceptance and maintenance of IC equipment. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/nci732Additional details
Identifiers
- DOI
- 10.1093/rpd/nci732;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 117
- Journal Issue
- 1-3
- Journal Page Range
- p. 79-86
- ISSN
- 0144-8420
Conference
- Title
- 3. Conference on Optimisation of dose and performance in interventional and digital imaging
- Dates
- 25-27 Mar 2004
- Place
- Leuven (Belgium)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 44036109
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- DESIGN; DOSE-RESPONSE RELATIONSHIPS; FLUOROSCOPY; HOSPITALS; IMAGE PROCESSING; PERFORMANCE; QUALITY CONTROL; RADIATION DOSES; RADIOCARDIOGRAPHY; RECOMMENDATIONS; STANDARDIZATION; STANDARDS; TECHNOLOGY ASSESSMENT
- Descriptors DEC
- BIOMEDICAL RADIOGRAPHY; BUILDINGS; CARDIOGRAPHY; CONTROL; DIAGNOSTIC TECHNIQUES; DOSES; MEDICAL ESTABLISHMENTS; MEDICINE; NUCLEAR MEDICINE; PROCESSING; RADIOLOGY
Optional Information
- Notes
- 30 refs