Published September 1997 | Version v1
Report Open

Development of beta reference radiations

Description

A system of beta reference radiation has been developed, that is composed of 740 MBq 147Pm beta source, 74 MBq and 740 MBq 90Sr + 90Y β sources, compensation filters, a source handling tool, a source jig, spacing bars, a shutter, a control unit and a beta dose meter calibration stand. For 740 MBq 147Pm and 74 MBq 90Sr + 90Y beta reference radiations with compensation filters and 740 MBq 90Sr + 90Y beta reference radiation without compensation filter, at 20 cm, 30 cm and 30 cm distance separately; the residual energy of maximum is 0.14 MeV, 1.98 MeV and 2.18 MeV separately; the absorbed dose to tissue D (0.07) is 1.547 mGy/h (1996-05-20), 5.037 mGy/h (1996-05-10) and 93.57 mGy/h (1996-05-15) separately; the total uncertainty is 3.0%, 1.7% and 1.7% separately. For the first and the second beta reference radiation, the dose rate variability in the area of 18 cm diameter in the plane perpendicular to the beta-ray beam axis is within +-6% and +-3% separately. (3 refs., 2 tabs., 8 figs.)

Availability note (English)

MF available from INIS under the Report Number.

Files

28074808.pdf

Files (228.2 kB)

Name Size Download all
md5:8890d0da338d56c0040576985270036f
228.2 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
8 p.
Report number
CNIC--01192

Optional Information

Secondary number(s)
IAE--0173.