Validation of a virtual source model of medical linac for Monte Carlo dose calculation using multi-threaded Geant4
Creators
- 1. Faculty of Science, Mohammed V University in Rabat, Rabat (Morocco)
Description
The use of phase space in medical linear accelerator Monte Carlo (MC) simulations significantly improves the execution time and leads to results comparable to those obtained from full calculations. The classical representation of phase space stores directly the information of millions of particles, producing bulky files. This paper presents a virtual source model (VSM) based on a reconstruction algorithm, taking as input a compressed file of roughly 800 kb derived from phase space data freely available in the International Atomic Energy Agency (IAEA) database. This VSM includes two main components; primary and scattered particle sources, with a specific reconstruction method developed for each. Energy spectra and other relevant variables were extracted from IAEA phase space and stored in the input description data file for both sources. The VSM was validated for three photon beams: Elekta Precise 6 MV/10 MV and a Varian TrueBeam 6 MV. Extensive calculations in water and comparisons between dose distributions of the VSM and IAEA phase space were performed to estimate the VSM precision. The Geant4 MC toolkit in multi-threaded mode (Geant4-[mt]) was used for fast dose calculations and optimized memory use. Four field configurations were chosen for dose calculation validation to test field size and symmetry effects, , , and for squared fields, and for an asymmetric rectangular field. Good agreement in terms of formalism, for 3%/3 mm and 2%/3 mm criteria, for each evaluated radiation field and photon beam was obtained within a computation time of 60 h on a single WorkStation for a 3 mm voxel matrix. Analyzing the VSM's precision in high dose gradient regions, using the distance to agreement concept (DTA), showed also satisfactory results. In all investigated cases, the mean DTA was less than 1 mm in build-up and penumbra regions. In regards to calculation efficiency, the event processing speed is six times faster using Geant4-[mt] compared to sequential Geant4, when running the same simulation code for both. The developed VSM for 6 MV/10 MV beams widely used, is a general concept easy to adapt in order to reconstruct comparable beam qualities for various linac configurations, facilitating its integration for MC treatment planning purposes. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1361-6560/aab7a1Additional details
Identifiers
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 63
- Journal Issue
- 8
- Journal Page Range
- [15 p.]
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 52002876
- Subject category
- S43: PARTICLE ACCELERATORS; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; COMMERCIAL BUILDINGS; DIFFERENTIAL THERMAL ANALYSIS; ENERGY SPECTRA; IAEA; LINEAR ACCELERATORS; MONTE CARLO METHOD; PHASE SPACE; PHOTON BEAMS; RADIATION DOSE DISTRIBUTIONS; SIMULATION; VALIDATION; VIBRATING SAMPLE MAGNETOMETERS
- Descriptors DEC
- ACCELERATORS; BEAMS; BUILDINGS; CALCULATION METHODS; INTERNATIONAL ORGANIZATIONS; MAGNETOMETERS; MATHEMATICAL SPACE; MEASURING INSTRUMENTS; SPACE; SPECTRA; TESTING; THERMAL ANALYSIS