A neutron spectrometer based on measurements of the time-of-flight of DD-neutrons over a 55 cm base has been developed. The paper gives detailed consideration to the circuit of a low-noise current preamplifier which ensures a coefficient of amplification of 2000 and a front rise time of 5 nsec. The stabilization of the transistor stages is achieved by incorporating a deep negative d.c. feedback. Space-time selection is accomplished by the method of electron collimation with coupling to the accompanying particles (helions)
Additional details
Additional titles
- Original title (Russian)
- Время-пролетный спектрометр в экспериментах с DD-нейтронами
Publishing Information
- Journal Title
- Prib. Tekh. Ehksp.
- Journal Issue
- no.4
- Series
- Prib. Tekh. Ehksp.
INIS
- Country of Publication
- USSR
- Country of Input or Organization
- USSR
- INIS RN
- 9379855
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BACKGROUND NOISE; DIAGRAMS; ENERGY RESOLUTION; FEEDBACK; NEUTRON SPECTRA; PERFORMANCE; PREAMPLIFIERS; TIME-OF-FLIGHT SPECTROMETERS
- Descriptors DEC
- AMPLIFIERS; ELECTRONIC EQUIPMENT; INFORMATION; MEASURING INSTRUMENTS; NOISE; RESOLUTION; SPECTRA; SPECTROMETERS
Optional Information
- Notes
- For English translation see the journal Instrum. Exp. Tech.