Published November 2001 | Version v1
Journal article

Electron dosimetry based on the absorbed dose to water concept: A comparison of the AAPM TG-51 and DIN 6800-2 protocols

  • 1. Abteilung Strahlentherapie, Universitaetsklinikum Freiburg, Hugstetter Strasse 55, 79106 Freiburg i. Br., (Germany)
  • 2. PTW Freiburg, Loerracher Strasse 7, 79115 Freiburg i. Br., (Germany)
  • 3. Abteilung fuer Medizinische Physik, Universitaetsklinik fuer Radioonkologie, Hoppe-Seyler-Str. 3, 72076 Tuebingen, (Germany)

Description

The dosimetry protocols DIN 6800-2 and AAPM TG-51, both based on the absorbed dose to water concept, are compared in their theoretical background and in their application to electron dosimetry. The agreement and disagreement in correction factors and energy parameters used in both protocols will be shown and discussed. Measurements with three different types of ionization chambers were performed and evaluated according to both protocols. As a result the perturbation correction factor Pwall60Co for the Roos chamber was determined to 1.024 ± 0.5%

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
28
Journal Issue
11
Journal Page Range
p. 2258-2264
ISSN
0094-2405
CODEN
MPHYA6

Optional Information

Notes
(c) 2001 American Association of Physicists in Medicine.