Published November 1996 | Version v1
Journal article

MR imaging of articular cartilage in the ankle: comparison of available imaging sequences and methods of measurement in cadavers

  • 1. Department of Radiology, Chang Gung Memorial Hospital, Linkou Medical Center, Taoyuan, Taiwan (Taiwan)
  • 2. University of California Medical Center, San Diego, CA (United States)
  • 3. Department of Radiology, Veterans Administrative Medical Center, San Diego, CA (United States)
  • 4. Department of Radiology, Veterans General Hospital-Taipei (Taiwan)

Description

Objective. To assess hyaline cartilage of cadaveric ankles using different magnetic resonance (MR) imaging techniques and various methods of measurement. Design and patients. Cartilage thicknesses of the talus and tibia were measured in ten cadaveric ankles by naked eye and by digitized image analysis from MR images of fat-suppressed T1-weighted gradient recalled (FS-SPGR), sequences and pulsed transfer saturation sequences with (FS-STS) and without fat-suppression (STS); these measurements were compared with those derived from direct inspection of cadaveric sections. The accuracy and precision errors were evaluated statistically for each imaging technique as well as measuring method. Contrast-to-noise ratios of cartilage versus joint fluid and marrow were compared for each of the imaging sequences. Results. Statistically, measurements from FS-SPGR images were associated with the smallest estimation error. Precision error of measurements derived from digitized image analysis was found to be smaller than that derived from naked eye measurements. Cartilage thickness measurements in images from STS and FS-STS sequences revealed larger errors in both accuracy and precision. Interobserver variance was larger in naked eye assessment of the cartilage. Contrast-to-noise ratio of cartilage versus joint fluid and marrow was higher with FS-SPGR than with FS-STS or STS sequences. Conclusion. Of the sequences and measurement techniques studied, the FS-SPGR sequence combined with the use of digitized image analysis provides the most accurate method for the assessment of ankle hyaline cartilage. (orig.). With 3 figs., 2 tabs

Additional details

Publishing Information

Journal Title
Skeletal Radiology
Journal Volume
25
Journal Issue
8
Journal Page Range
p. 749-755.
ISSN
0364-2348
CODEN
SKRADI

INIS

Country of Publication
Germany
Country of Input or Organization
Germany
INIS RN
28018098
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ANATOMY; AUTOPSY; BONE JOINTS; CARTILAGE; DIAGNOSTIC TECHNIQUES; FEET; NMR IMAGING; SKELETON
Descriptors DEC
ANIMAL TISSUES; BIOLOGY; BODY; BODY AREAS; CONNECTIVE TISSUE; LEGS; LIMBS; ORGANS