Published January 12, 2015 | Version v1
Journal article

Current status of 3D EPID-based in vivo dosimetry in The Netherlands Cancer Institute

  • 1. Department of Radiation Oncology, The Netherlands Cancer Institute, Plesmanlaan 121, 1066 CX Amsterdam (Netherlands)

Description

3D in vivo dose verification using a-Si EPIDs is performed routinely in our institution for almost all RT treatments. The EPID-based 3D dose distribution is reconstructed using a back-projection algorithm and compared with the planned dose distribution using 3D gamma evaluation. Dose-reconstruction and gamma-evaluation software runs automatically, and deviations outside the alert criteria are immediately available and investigated, in combination with inspection of cone-beam CT scans. The implementation of our 3D EPID- based in vivo dosimetry approach was able to replace pre-treatment verification for more than 90% of the patient treatments. Clinically relevant deviations could be detected for approximately 1 out of 300 patient treatments (IMRT and VMAT). Most of these errors were patient related anatomical changes or deviations from the routine clinical procedure, and would not have been detected by pre-treatment verification. Moreover, 3D EPID-based in vivo dose verification is a fast and accurate tool to assure the safe delivery of RT treatments. It provides clinically more useful information and is less time consuming than pre-treatment verification measurements. Automated 3D in vivo dosimetry is therefore a prerequisite for large-scale implementation of patient-specific quality assurance of RT treatments

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/573/1/012014

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
573
Journal Issue
1
Journal Page Range
[4 p.]
ISSN
1742-6596

Conference

Title
8. International Conference on 3D Radiation Dosimetry
Acronym
IC3DDose
Dates
4-7 Sep 2014
Place
Ystad (Sweden)

INIS