TU-CD-304-10: Development and Optimization of "Compton Lens" Collimator Design for Increased Dose Rate in SRS

  • 1. The University of Wisconsin-Madison, Madison, WI (United States)

Description

Purpose: To provide a proof of concept for a new collimator design to increase the dose rate at isocenter for stereotactic radiosurgery (SRS) treatments by taking advantage of off axis Compton scattered photons which are attenuated in current collimators. Methods: A fundamentally new collimator design was developed and optimized by introducing a series of slits to a standard block collimator. The introduction of slits allowed for initially off axis radiation that was scattered in the direction of the target to contribute to the target dose. For optimization, the design was broken into two parts: an upper interaction plate where primary scattering occurs, and a lower "Compton slit" region which allows for scattered photons traveling toward the target to reach isocenter. To optimize the design, a series of simulations were performed using MCNP6 in which several key parameters were adjusted and the output was compared to a standard collimator. Key parameters modified included the collimator material, cone size, and interaction plate thickness. The effects of using energies different than the traditional 6 MV beam were also explored. Results: An optimized collimator design utilizing a solid interaction plate with a Cesium-137 beam and a 4 mm cone size resulted in a dose rate increase on the order of 5% relative to standard collimators in use. Currently, designs incorporating a Cesium-137 source are the most feasible due to necessary size and weight concerns for 6 MV beams. Conclusion: Preliminary designs provide a proof of concept and indicate a potential to improve upon the dose rate of current collimators while not largely compromising the sharp dose falloff inherent to SRS. Further optimization into the geometry and positioning of the interaction plate, as well as slit optimization, will likely lead to further dose rate increases than were observed in this study. Funding for this work was provided by the Wisconsin Alumni Research Foundation (WARF). Authors have a conflict of interest as the technology discussed is patent pending, and thus there is a financial interest in the work

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
42
Journal Issue
6
Journal Page Range
vp.
ISSN
0094-2405
CODEN
MPHYA6

Optional Information

Notes
(c) 2015 American Association of Physicists in Medicine