Image information transfer properties of x-ray fluorescent screens
Description
The image information transfer efficiency for five x-ray fluorescent screens (calcium tungstate, barium halide, and three rare earth screens) has been experimentally determined with monoenergetic x-ray beams at energies of 18, 22, 32, 49, 51, 58, and 69 keV. The transfer efficiency, which is defined by the ratio of the output signal-to-noise ratios, was determined from measurements of (a) the fraction of incident x rays absorbed in the screen and (b) the statistical distribution of the number of light photons emitted from the screen per absorbed x ray which was determined by light photon counting techniques. Comparisons of the information transfer efficiency, the average number of light photons emitted per absorbed x ray, and the light output energy per Roentgen are given for the above screens and x-ray energies
Additional details
Publishing Information
- Journal Title
- Med. Phys.
- Journal Volume
- 8
- Journal Issue
- 3
- Series
- Med. Phys.
- Journal Page Range
- 337-46
- ISSN
- 0094-2405
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 14719699
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- EFFICIENCY; SCREENS; X RADIATION; X-RAY FLUORESCENCE ANALYSIS
- Descriptors DEC
- CHEMICAL ANALYSIS; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; NONDESTRUCTIVE ANALYSIS; RADIATIONS; X-RAY EMISSION ANALYSIS