Published May 1981 | Version v1
Journal article

Image information transfer properties of x-ray fluorescent screens

  • 1. National Bureau of Standards, Washington, D.C.

Description

The image information transfer efficiency for five x-ray fluorescent screens (calcium tungstate, barium halide, and three rare earth screens) has been experimentally determined with monoenergetic x-ray beams at energies of 18, 22, 32, 49, 51, 58, and 69 keV. The transfer efficiency, which is defined by the ratio of the output signal-to-noise ratios, was determined from measurements of (a) the fraction of incident x rays absorbed in the screen and (b) the statistical distribution of the number of light photons emitted from the screen per absorbed x ray which was determined by light photon counting techniques. Comparisons of the information transfer efficiency, the average number of light photons emitted per absorbed x ray, and the light output energy per Roentgen are given for the above screens and x-ray energies

Additional details

Publishing Information

Journal Title
Med. Phys.
Journal Volume
8
Journal Issue
3
Series
Med. Phys.
Journal Page Range
337-46
ISSN
0094-2405

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
14719699
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
EFFICIENCY; SCREENS; X RADIATION; X-RAY FLUORESCENCE ANALYSIS
Descriptors DEC
CHEMICAL ANALYSIS; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; NONDESTRUCTIVE ANALYSIS; RADIATIONS; X-RAY EMISSION ANALYSIS