Published October 1, 2006 | Version v1
Journal article

Improvement of sensitivity of multisample biological immunoassay system using HTS SQUID and magnetic nanoparticles

  • 1. Advanced Research Laboratory, Hitachi Ltd., 1-280, Higashi-Koigakubo, Kokubunji-shi, Tokyo 185-8601 (Japan)
  • 2. Central Research Laboratory, Hitachi Ltd., Kokubunji-shi, Tokyo 185-8601 (Japan)
  • 3. Department of Clinical Chemistry and Laboratory Medicine, Kyushu University, Fukuoka 812-8582 (Japan)
  • 4. Plastic Products Division, INOAC Corporation, Nagoya 456-0054 (Japan)
  • 5. Department of Marketing, Kyushu INOAC, Fukuoka 812-0013 (Japan)
  • 6. Research Institute of Superconductor Science and Systems, Kyushu University, Fukuoka 812-8581 (Japan)

Description

Recently, we have developed a prototype magnetic immunoassay system using a high temperature superconductor (HTS) superconducting quantum interference device (SQUID) to investigate the performance and usability of the magnetic immunoassay. In this study, we improved the immunoassay system to heighten the sensitivity of the immunoassay measurement. To reduce the SQUID-to-sample distance, we introduced a structure to compensate for thermal shrinkage in the cryostat and reduce the warpage of the window. The shape of pickup coil was also optimized to improve the field sensitivity. After these improvements, the magnetic signal intensity from Fe3O4 nanoparticles became about two times stronger than that achieved by our original system

Additional details

Identifiers

DOI
10.1016/j.physc.2006.06.019;
PII
S0921-4534(06)00615-0;

Publishing Information

Journal Title
Physica. C, Superconductivity
Journal Volume
445-448
Journal Page Range
p. 975-978
ISSN
0921-4534
CODEN
PHYCE6

Conference

Title
18. international symposium on superconductivity
Acronym
ISS 2005
Dates
24-25 Oct 2005
Place
Tsukuba (Japan)

Optional Information

Copyright
Copyright (c) 2006 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.