Published October 21, 2011 | Version v1
Journal article

Direct impact analysis of multi-leaf collimator leaf position errors on dose distributions in volumetric modulated arc therapy: a pass rate calculation between measured planar doses with and without the position errors

  • 1. Department of Radiology, Osaka City University Hospital, Osaka (Japan)
  • 2. Department of Radiology, Osaka City University Graduate School of Medicine, Osaka (Japan)

Description

We propose a new method for analyzing the direct impact of multi-leaf collimator (MLC) leaf position errors on dose distributions in volumetric modulated arc therapy (VMAT). The technique makes use of the following processes. Systematic leaf position errors are generated by directly changing a leaf offset in a linac controller; dose distributions are measured by a two-dimensional diode array; pass rates of the dose difference between measured planar doses with and without the position errors are calculated as a function of the leaf position error. Three different treatment planning systems (TPSs) were employed to create VMAT plans for five prostate cancer cases and the pass rates were compared between the TPSs under various leaf position errors. The impact of the leaf position errors on dose distributions depended upon the final optimization result from each TPS, which was explained by the correlation between the dose error and the average leaf gap width. The presented method determines leaf position tolerances for VMAT delivery for each TPS, which may facilitate establishing a VMAT quality assurance program in a radiotherapy facility. (note)

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/56/20/N03

Additional details

Identifiers

DOI
10.1088/0031-9155/56/20/N03;
PII
S0031-9155(11)95739-X;

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
56
Journal Issue
20
Journal Page Range
p. N237-N246
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
43022511
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COLLIMATORS; DOSES; ERRORS; NEOPLASMS; OPTIMIZATION; PLANNING; PROSTATE; QUALITY ASSURANCE; RADIATION DOSE DISTRIBUTIONS; RADIOTHERAPY
Descriptors DEC
BODY; DISEASES; GLANDS; MALE GENITALS; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; THERAPY