Long-term accuracy and precision of PIXE and PIGE measurements for thin and thick sample analyses
Description
This paper describes PIXE/PIGE measurements on thin Micromatter Standard (±5%) foils run over a period of 10 years. The selected foils were typically 50 μg/cm2 thick and covered the commonly used PIXE X-ray energy range 1.4-20 keV and the light elements F and Na for PIGE studies. For the thousands of thick obsidian and pottery samples analysed over a 6-year period, the Ohio Red Clay standard has been used for both PIXE and PIGE calibration of a range of elements from Li to Rb. For PIXE, the long-term accuracy could be as low as ±1.6% for major elements with precision ranging from ±5% to ±10% depending on the elemental concentration. For PIGE, accuracies were around ±5% with precision ranging from ±5% in thick samples to ±15% in thin samples or for low yield γ-ray production
Additional details
Identifiers
- PII
- S0168583X01010114;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section B, Beam Interactions with Materials and Atoms
- Journal Volume
- 189
- Journal Issue
- 1-4
- Journal Page Range
- p. 81-85
- ISSN
- 0168-583X
- CODEN
- NIMBEU
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 33059315
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACCURACY; CLAYS; KEV RANGE; NUCLEAR REACTION ANALYSIS; PIXE ANALYSIS; PROMPT GAMMA RADIATION
- Descriptors DEC
- CHEMICAL ANALYSIS; ELECTROMAGNETIC RADIATION; ENERGY RANGE; GAMMA RADIATION; IONIZING RADIATIONS; MINERALS; NONDESTRUCTIVE ANALYSIS; RADIATIONS; SILICATE MINERALS; X-RAY EMISSION ANALYSIS
Optional Information
- Copyright
- Copyright (c) 2002 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.