Body potassium measurement using Phoswich based lung monitor

  • 1. Radiation Safety Section, Indira Gandhi Centre for Atomic Research, Kalpakkam (India)

Description

Phoswich based lung monitor enabled with the feature of high energy photon measurement is effectively utilised to measure body potassium of radiation workers reported for lung monitoring at RSD, IGCAR. The Phoswich systems, kept inside the low background steel room in an optimum counting geometry, has resulted in a higher efficiency value and lesser MDA for 40K compared to whole body counters (4 inch x 3 inch NaI(Tl) - scanning geometry) due to increased counting time and reduced background in steel room. This paper discusses about calibration, body potassium measurement, its correlation with parameters like gender and the results obtained

Part of:
Proceedings of the international conference on radiological safety in workplace, nuclear facilities and environment: book of abstracts

Additional details

Publishing Information

Publisher
Indian Association for Radiation Protection
Imprint Place
Mumbai (India)
Imprint Title
Proceedings of the international conference on radiological safety in workplace, nuclear facilities and environment: book of abstracts
Imprint Pagination
287 p.
Journal Page Range
p. 114

Conference

Title
32. international conference on radiological safety in workplace, nuclear facilities and environment
Acronym
IARPIC-2016
Dates
22-25 Feb 2016
Place
Kalpakkam (India)

Optional Information

Notes
1 ref., 2 figs., 1 tab.