Development of a Si-PM-based high-resolution PET system for small animals
Creators
- 1. Kobe City College of Technology, Kobe (Japan)
- 2. Department of Nuclear Medicine and Tracer Kinetics, Osaka University Graduate School of Medicine, Osaka (Japan)
- 3. Department of Molecular Imaging in Medicine, Osaka University Graduate School of Medicine, Osaka (Japan)
Description
A Geiger-mode avalanche photodiode (Si-PM) is a promising photodetector for PET, especially for use in a magnetic resonance imaging (MRI) system, because it has high gain and is less sensitive to a static magnetic field. We developed a Si-PM-based depth-of-interaction (DOI) PET system for small animals. Hamamatsu 4 x 4 Si-PM arrays (S11065-025P) were used for its detector blocks. Two types of LGSO scintillator of 0.75 mol% Ce (decay time: ∼45 ns; 1.1 mm x 1.2 mm x 5 mm) and 0.025 mol% Ce (decay time: ∼31 ns; 1.1 mm x 1.2 mm x 6 mm) were optically coupled in the DOI direction to form a DOI detector, arranged in a 11 x 9 matrix, and optically coupled to the Si-PM array. Pulse shape analysis was used for the DOI detection of these two types of LGSOs. Sixteen detector blocks were arranged in a 68 mm diameter ring to form the PET system. Spatial resolution was 1.6 mm FWHM and sensitivity was 0.6% at the center of the field of view. High-resolution mouse and rat images were successfully obtained using the PET system. We confirmed that the developed Si-PM-based PET system is promising for molecular imaging research.
Availability note (English)
Available from http://dx.doi.org/10.1088/0031-9155/55/19/013Additional details
Identifiers
- DOI
- 10.1088/0031-9155/55/19/013;
- PII
- S0031-9155(10)60352-1;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 55
- Journal Issue
- 19
- Journal Page Range
- p. 5817-5831
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 42035825
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- MICE; NMR IMAGING; PHOTODIODES; POSITRON COMPUTED TOMOGRAPHY; PULSE SHAPERS; RATS; SCINTILLATOR-PHOTODIODE DETECTORS
- Descriptors DEC
- ANIMALS; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; ELECTRONIC CIRCUITS; EMISSION COMPUTED TOMOGRAPHY; MAMMALS; MEASURING INSTRUMENTS; PULSE CIRCUITS; RADIATION DETECTORS; RODENTS; SCINTILLATION COUNTERS; SEMICONDUCTOR DEVICES; SEMICONDUCTOR DIODES; SIGNAL CONDITIONERS; TOMOGRAPHY; VERTEBRATES