Published January 1990 | Version v1
Journal article

Spot scanning for 250 MeV protons

  • 1. Paul Scherrer Inst., Villigen (Switzerland). Abt. Strahlenmedizin

Description

At the Swiss Institute for Nuclear Research a computer optimized treatment planning system has been developed for pion radiotherapy. The principle of this dynamic treatment technique can be adapted for proton therapy. It could prove to be a valid alternative to passive scattering for proton treatment especially of large irregularly shaped volumes. Dose calculations have been performed for specific cases using a 250 MeV pencil beam. The dose distributions and dose volume histograms have been compared with pion therapy. A horizontal experimental beam line, using a degraded 590 MeV proton beam of the PSI ring cyclotron, is set up to verify calculations in phantoms. (orig./MG)

Additional details

Publishing Information

Journal Title
Strahlentherapie und Onkologie
Journal Volume
166
Journal Issue
1
Series
Strahlenther. Onkol.
Journal Page Range
45-48
ISSN
0179-7158
CODEN
STONE

Conference

Title
International workshop on particle therapy.
Dates
28-29 Oct 1988.
Place
Essen (Germany, F.R.).