Published January 1990
| Version v1
Journal article
Spot scanning for 250 MeV protons
Creators
- 1. Paul Scherrer Inst., Villigen (Switzerland). Abt. Strahlenmedizin
Description
At the Swiss Institute for Nuclear Research a computer optimized treatment planning system has been developed for pion radiotherapy. The principle of this dynamic treatment technique can be adapted for proton therapy. It could prove to be a valid alternative to passive scattering for proton treatment especially of large irregularly shaped volumes. Dose calculations have been performed for specific cases using a 250 MeV pencil beam. The dose distributions and dose volume histograms have been compared with pion therapy. A horizontal experimental beam line, using a degraded 590 MeV proton beam of the PSI ring cyclotron, is set up to verify calculations in phantoms. (orig./MG)
Additional details
Publishing Information
- Journal Title
- Strahlentherapie und Onkologie
- Journal Volume
- 166
- Journal Issue
- 1
- Series
- Strahlenther. Onkol.
- Journal Page Range
- 45-48
- ISSN
- 0179-7158
- CODEN
- STONE
Conference
- Title
- International workshop on particle therapy.
- Dates
- 28-29 Oct 1988.
- Place
- Essen (Germany, F.R.).
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 21031979
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- COMPUTER CALCULATIONS; CYCLOTRONS; MEV RANGE 100-1000; OPTIMIZATION; PHANTOMS; PLANNING; PROTON BEAMS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY; SCATTERING; VOLUME
- Descriptors DEC
- ACCELERATORS; BEAMS; CYCLIC ACCELERATORS; ENERGY RANGE; MEDICINE; MEV RANGE; MOCKUP; NUCLEON BEAMS; PARTICLE BEAMS; STRUCTURAL MODELS; THERAPY