Published March 1996 | Version v1
Journal article

The Texas Thermal Interface: A real-time computer interface for an Inframetrics infrared camera

  • 1. Fusion Research Center, University of Texas, Austin, Texas 78712 (United States)

Description

The Texas Thermal Interface (TTI) offers an advantageous alternative to the conventional video path for computer analysis of infrared images from Inframetrics cameras. The TTI provides real-time computer data acquisition of 48 consecutive fields (version described here) with 8-bit pixels. The alternative requires time-consuming individual frame grabs from video tape with frequent loss of resolution in the D/A/D conversion. Within seconds after the event, the TTI temperature files may be viewed and processed to infer heat fluxes or other quantities as needed. The system cost is far less than commercial units which offer less capability. The system was developed for and is being used to measure heat fluxes to the plasma-facing components in a tokamak. copyright 1996 American Institute of Physics

Additional details

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
67
Journal Issue
3
Journal Page Range
p. 726-729.
ISSN
0034-6748
CODEN
RSINAK