Implementing Dosimetry in the Clinic. Chapter 11
Description
Clinical dosimetry is a multistep procedure (see Fig. 11.1); each step is essential to obtain an accurate estimate of the absorbed dose delivered to one or more tumours or normal tissues. The first step in the clinical dosimetry workflow is the calibration procedure (step 1). This is a prerequisite for activity determination and may require phantom image acquisition with a known amount of activity that should ideally be traceable to a standard laboratory. Procedures for obtaining quantitative information on the activity or activity concentration from images were presented in Chapter 4 and will include sensitivity determination and may include partial volume and dead time corrections. The calibration procedure is a key step in accurately determining the activity distribution. The calibration step and patient image/data acquisition step (step 2) are linked; the camera settings used in the quantitative imaging calibration procedure should be used for all patient imaging. For single photon emission computed tomography (SPECT) quantitative imaging this may mean that the acquisition matrix size, the type of collimator or the number and position of energy windows should remain constant between phantom calibration and patient imaging.
Additional details
Identifiers
Publishing Information
- Publisher
- IAEA
- Imprint Place
- Vienna (International Atomic Energy Agency (IAEA))
- ISBN
- 978-92-0-139523-8
- Imprint Title
- Dosimetry for Radiopharmaceutical Therapy
- Imprint Pagination
- 320 p.
- Journal Page Range
- p. 278-294
INIS
- Country of Publication
- International Atomic Energy Agency (IAEA)
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- CALIBRATION; CALIBRATION STANDARDS; CAMERAS; COLLIMATORS; CORRECTIONS; DATA ACQUISITION; DOSIMETRY; GAMMA CAMERAS; IMAGE PROCESSING; IMAGES; NEOPLASMS; PATIENTS; PHANTOMS; RADIATION DOSE DISTRIBUTIONS; SENSITIVITY; SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY
- Descriptors DEC
- CAMERAS; COMPUTERIZED TOMOGRAPHY; DATA PROCESSING; DIAGNOSTIC TECHNIQUES; DISEASES; EMISSION COMPUTED TOMOGRAPHY; MOCKUP; PROCESSING; STANDARDS; STRUCTURAL MODELS; TOMOGRAPHY
Optional Information
- Notes
- 36 refs., 1 tab., 3 figs.
- Secondary number(s)
- STI/PUB--2065