Published May 1, 2008 | Version v1
Journal article

Multileaf Collimator Characteristics and Reliability Requirements for IMRT Elekta System

  • 1. Department of Radiation Oncology, University of Florida College of Medicine, Gainesville, FL (United States)
  • 2. Sun Nuclear Corporation, Melbourne, FL (Australia)

Description

Understanding the characteristics of a multileaf collimator (MLC) system, modeling MLC in a treatment planning system, and maintaining the mechanical accuracy of the linear accelerator gantry head system are important factors in the safe implementation of an intensity-modulated radiotherapy program. We review the characteristics of an Elekta MLC system, discuss the necessary MLC modeling parameters for a treatment planning system, and provide a novel method to establish an MLC leaf position quality assurance program. To perform quality assurance on 40 pairs of individual MLC leaves is a time-consuming and difficult task. In this report, an effective routine MLC quality assurance method based on the field edge of a backup jaw as referenced in conjunction with a diode array as a radiation detector system is discussed. The sensitivity of this test for determining the relative leaf positions was observed to be better than 0.1 mm. The Elekta MLC leaf position accuracy measured with this system has been better than 0.3 mm

Availability note (English)

Available from http://dx.doi.org/10.1016/j.ijrobp.2007.07.2392

Additional details

Identifiers

DOI
10.1016/j.ijrobp.2007.07.2392;
PII
S0360-3016(07)04354-4;

Publishing Information

Journal Title
International Journal of Radiation Oncology, Biology and Physics
Journal Volume
71
Journal Issue
1
Journal Page Range
p. S89-S92
ISSN
0360-3016
CODEN
IOBPD3

Conference

Title
Challenges of advanced technology
Acronym
2007 interorganizational symposium on quality assurance of radiation therapy
Dates
20-22 Feb 2007
Place
Dallas, TX (United States)

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
40006936
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ACCURACY; COLLIMATORS; LINEAR ACCELERATORS; QUALITY ASSURANCE; RADIATION DETECTORS; RADIOTHERAPY; RELIABILITY; REVIEWS; SIMULATION
Descriptors DEC
ACCELERATORS; DOCUMENT TYPES; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY

Optional Information

Copyright
Copyright (c) 2008 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.