Published February 2011 | Version v1
Report

Relative calibration of SXR detector array

  • 1. Univ. of Science and Technology of China, Hefei, Anhui (China)
  • 2. Inst. of Plasma Physics, Chinese Academy of Sciences, Hefei, Anhui (China)

Description

A calibration system of SXR detectors was established. The system employed a standard luminance source which is working in visible spectral instead of a soft- x-ray source. With this system we studied the detector arrays which were used four experimental campaigns on EAST tokamak in previous experiments. Reduction of spectrum averaged efficiency of the array had been observed. Comparing results with the new detector, we also found aging progress could improve the coherency of the array. (author)

Part of:
Proceeding of JSPS-CAS Core University Program seminar on production and control of high performance plasmas with advanced plasma heating and diagnostic systems

Additional details

Publishing Information

Imprint Title
Proceeding of JSPS-CAS Core University Program seminar on production and control of high performance plasmas with advanced plasma heating and diagnostic systems
Imprint Pagination
385 p.
Journal Page Range
p. 235-241
Report number
NIFS-PROC--84

Conference

Title
JSPS-CAS Core University Program (CUP) seminar on production and control of high performance plasmas with advanced plasma heating and diagnostic systems
Dates
1-4 Nov 2010
Place
Guilin (China)

Optional Information

Notes
5 refs., 2 figs.