Published November 2014 | Version v1
Journal article

Effective usage of a clearance check to avoid a collision in Gamma Knife Perfexion radiosurgery with the Leksell skull frame

  • 1. Nagoya University, Graduate School of Medicine, Department of Radiological Sciences, Nagoya, Aichi (Japan)
  • 2. Nagoya Kyoritsu Hospital, Nagoya Radiosurgery Center, Nagoya, Aichi (Japan)
  • 3. Aichi Medical University, Department of Radiology and Radiation Oncology, Nagakute, Aichi (Japan)
  • 4. Nagoya City University, Graduate School of Medical Sciences, Department of Radiology, Nagoya, Aichi (Japan)

Description

Skull frame attachment is one of the most significant issues with Gamma Knife radiosurgery. Because of the potential for suffering by patients, careful control of the frame position is required to avoid circumstances such as collision between the frame or the patient's head and the collimator helmet, and inaccessible target coordinates. This study sought to develop a simulation method to find the appropriate frame location on the patient's head by retrospective analysis of treatment plans for brain metastasis cases. To validate the accuracy of the collision warning, we compared the collision distance calculated using Leksell Gamma Plan (LGP) with actual measured distances. We then investigated isocenter coordinates in near-collision cases using data from 844 previously treated patients and created a clearance map by superimposing them on CT images for just the frame, post and stereotactic fiducial box. The differences in distance between the simulation in LGP and the measured values were <1.0 mm. In 177 patients, 213 lesions and 461 isocenters, there was a warning of one possible collision. The clearance map was helpful for simulating appropriate skull frame placement. The clearance simulation eliminates the psychological stress associated with potential collisions, and enables more comfortable treatment for the patient. (author)

Availability note (English)

Available from http://dx.doi.org/10.1093/jrr/rru057

Additional details

Identifiers

Publishing Information

Journal Title
Journal of Radiation Research
Journal Volume
55
Journal Issue
6
Journal Page Range
p. 1192-1198
ISSN
0449-3060

Optional Information

Notes
7 refs., 6 figs., 1 tab.