Published 2005 | Version v1
Journal article

Phantom studies for possible dose reduction in CT head procedures

  • 1. Nofer Inst. of Occupational Medicine, Lodz (Poland)
  • 2. Regional Special Hospital, Rzeszow (Poland)
  • 3. Polish Mother's Memorial Hospital-Research Inst., Lodz (Poland)

Description

This paper presents the results of phantom studies to investigate possible dose reduction in relation to image quality in head examinations. The studies were performed using five single-slice computed tomography (CT) scanners. Beginning from the manufacturer's protocols (i.e. default protocols in the scanner software) for routine head (adult) examinations, the values of kVp, anode current and time were modified. Low-contrast resolution and spatial resolution were controlled using a Catphan 424 phantom. Radiation dose was checked using a polymethylmethacrylate phantom and a pencil ionisation chamber. It was found that CT dose index may be reduced up to three times with practically no loss of image quality. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/nch571

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
114
Journal Issue
1-3
Journal Page Range
p. 326-331
ISSN
0144-8420

Conference

Title
2. Malmoe Conference on Medical X-Ray Imaging - Optimisation strategies in medical X-ray imaging
Dates
23-25 Apr 2004
Place
Malmoe (Sweden)

Optional Information

Notes
5 refs