Phantom studies for possible dose reduction in CT head procedures
Creators
- 1. Nofer Inst. of Occupational Medicine, Lodz (Poland)
- 2. Regional Special Hospital, Rzeszow (Poland)
- 3. Polish Mother's Memorial Hospital-Research Inst., Lodz (Poland)
Description
This paper presents the results of phantom studies to investigate possible dose reduction in relation to image quality in head examinations. The studies were performed using five single-slice computed tomography (CT) scanners. Beginning from the manufacturer's protocols (i.e. default protocols in the scanner software) for routine head (adult) examinations, the values of kVp, anode current and time were modified. Low-contrast resolution and spatial resolution were controlled using a Catphan 424 phantom. Radiation dose was checked using a polymethylmethacrylate phantom and a pencil ionisation chamber. It was found that CT dose index may be reduced up to three times with practically no loss of image quality. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/nch571Additional details
Identifiers
- DOI
- 10.1093/rpd/nch571;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 114
- Journal Issue
- 1-3
- Journal Page Range
- p. 326-331
- ISSN
- 0144-8420
Conference
- Title
- 2. Malmoe Conference on Medical X-Ray Imaging - Optimisation strategies in medical X-ray imaging
- Dates
- 23-25 Apr 2004
- Place
- Malmoe (Sweden)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 44035816
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- COMPUTER CODES; COMPUTERIZED TOMOGRAPHY; HEAD; IMAGES; IONIZATION CHAMBERS; PHANTOMS; RADIATION DOSES; SPATIAL RESOLUTION; X RADIATION
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; DOSES; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEASURING INSTRUMENTS; MOCKUP; RADIATION DETECTORS; RADIATIONS; RESOLUTION; STRUCTURAL MODELS; TOMOGRAPHY
Optional Information
- Notes
- 5 refs