Development of radiation dose calculation software using the size-specific dose estimate
Creators
- 1. Department of Radiology, The Jikei University Hospital, 3-19-18 Nishi-Shimbashi, Minato-ku, Tokyo 105-8471 (Japan)
- 2. Advanced Imaging Center Yaesu Clinic, 2-1-18 Nihombashi, Chuo-ku, Tokyo 103-0027 (Japan)
- 3. Department of Diagnostic Imaging, The Jikei University Hospital, 3-19-18 Nishi-Shimbashi, Minato-ku, Tokyo 105-8471 (Japan)
- 4. Department of Radiology, Stony Brook University School of Medicine, 101 Nicolls Road Stony Brook, NY 11794 (United States)
- 5. Department of Radiology, Boston University School of Medicine, 820 Harrison Avenue, FGH Building 3rd Floor, Boston, MA 02118 (United States)
- 6. Department of Radiological Sciences, Tokyo Metropolitan University, 7-2-10 Higashi-Ogu, Arakawa-ku, Tokyo 116-8551 (Japan)
Description
We aimed to develop a software for facilitating absorbed dose per pixel (pixel dose) calculation using a size-specific dose estimate (SSDE). We calculated the pixel dose at nine equal points inserted into the radiophotoluminescence glass dosemeter (RPLD) and compared the pixel dose with the measured doses using RPLD. With this method, the relative errors of average pixel dose was -0.1% for adults and 2.86, 3.36 and 1.17% for those aged 10, 5 and 1 years without tube current modulation, respectively. In contrast, the relative error of SSDE was 17.37% for adults and 20.38, 20.73 and 19.20% for those aged 10, 5 and 1 years, respectively. In other words, the pixel dose was almost equal to the measured doses. Therefore, our software can be useful for determining arbitrary point. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncy074Additional details
Identifiers
- DOI
- 10.1093/rpd/ncy074;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 182
- Journal Issue
- 3
- Journal Page Range
- p. 359-369
- ISSN
- 0144-8420
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 50010890
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S97: MATHEMATICAL METHODS AND COMPUTING; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- ABSORBED RADIATION DOSES; COMPUTER CODES; ERRORS; MODULATION; RPL DOSEMETERS
- Descriptors DEC
- DOSEMETERS; DOSES; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; RADIATION DOSES
Optional Information
- Notes
- 24 refs.