Published 2014 | Version v1
Miscellaneous Open

Performance evaluation of activimeter for PET radiopharmaceutical measuring based on 18F

Description

Proficiency test is an essential tool to assess the reliability and accuracy of measurements. In this work, proficiency tests (Z-score, accuracy and relative deviation) were applied to evaluate the performance of the activimeter used at Radiopharmaceuticals Production Division of the Regional Center for Nuclear Sciences of the Northeast (CRCN-NE/CNEN). The results were evaluated in compliance with ISO/IEC Guide 43-1. Additionally, correction factors for different measurement geometries were determined experimentally for those devices. (author)

Files

46047004.pdf

Files (124.8 kB)

Name Size Download all
md5:5ee292dd9d9a2cf445a012210cf4f711
124.8 kB Preview Download

Additional details

Additional titles

Original title (Portuguese)
Avaliacao de desempenho de ativimetros para medicao de radiofarmacos PET baseados em 18F

Publishing Information

Imprint Pagination
4 p.
Report number
INIS-BR--15008

Conference

Title
Brazilian congress on metrology of ionizing radiation
Original Conference Title
CMRI 2014: Congresso brasileiro de metrologia das radiacoes ionizantes
Acronym
CMRI 2014
Dates
23-25 Nov 2014
Place