Published June 14, 1991 | Version v1
Journal article

Measurements of emission spectra during thermoluminescence (TL) from LiF(Mg, Cu, P) TL dosimeters

  • 1. Sussex Univ., Brighton (UK). School of Mathematical and Physical Sciences

Description

A detailed analysis of the spectrum of the light emitted during thermoluminescence (TL) from irradiated LiF (Mg, Cu, P) is presented. The data show how the emission characteristics vary as functions of the concentrations of the individual dopants (Mg, Cu and P) and of the pre-irradiation heat treatments. It is suggested that the data can be interpreted in terms of precipitated phases of Mg and P. It is speculated that the heat treatments cause the dissolution of these phases and this in turn alters the glow-curve structure and leads to a reduction of the TL sensitivity. (author)

Additional details

Publishing Information

Journal Title
Journal of Physics. D, Applied Physics
Journal Volume
24
Journal Issue
6
Series
J. Phys., D Appl. Phys.
Journal Page Range
988-996
ISSN
0022-3727
CODEN
JPAPB

Optional Information

Contract/Grant/Project number
Grant HR9-020/3720; 87-00101; INT-8821998