Published August 1987 | Version v1
Journal article

Improved external beam radiotherapy for the treatment of retinoblastoma

  • 1. Saint Bartholomew's Hospital, London (UK)
  • 2. Moorfields Hospital, London (UK)

Description

An applicator attached to a 6 MV linear accelerator has been adapted to produce a well defined beam for the treatment of retinoblastoma. By using a contact lens as the reference point on the patient's cornea, the beam edge is aligned to just behind the posterior margin of the crystalline lens, thus sparing the anterior chamber and lens from the radiation field. In addition to basic dosimetric measurements made using an ionisation chamber and film densitometry to quantify the output and dose distribution from the applicator, extensive use was made of thermoluminescent dosemeters and a tissue-equivalent anthropomorphic head phantom. This was designed for detailed dosimetry on the eye and was used to confirm the correct positioning and reproducibility of the clinical set-up. In addition to the improved dosimetric analysis, other advantages include patient immobilisation, anaesthetic procedure and finer couch control as well as a recommendation on the dose prescription. (author)

Additional details

Publishing Information

Journal Title
Br. J. Radiol.
Journal Volume
50
Journal Issue
716
Series
Br. J. Radiol.
Journal Page Range
753-760
ISSN
0007-1285
CODEN
BJRAA

INIS

Country of Publication
United Kingdom
Country of Input or Organization
United Kingdom
INIS RN
18095074
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
DOSIMETRY; ELECTRON BEAMS; MAN; NEOPLASMS; PARTIAL BODY IRRADIATION; RADIOTHERAPY; RETINA; SENSE ORGANS DISEASES
Descriptors DEC
ANIMALS; BEAMS; BODY; DISEASES; EXTERNAL IRRADIATION; EYES; IRRADIATION; LEPTON BEAMS; MAMMALS; MEDICINE; ORGANS; PARTICLE BEAMS; PRIMATES; SENSE ORGANS; THERAPY; VERTEBRATES