Published May 21, 2020 | Version v1
Journal article

Slice NEQ and system DQE to assess CT imaging performance

  • 1. Institute of Radiation Physics (IRA), Lausanne University Hospital (CHUV) and University of Lausanne, Rue du Grand-Pré 1, 1007 Lausanne (Switzerland)

Description

The standard Fourier metrics used for assessing image quality in computed tomography (CT) use a planar (2D) formulation of the noise equivalent quanta (NEQ) without a specific concept to manage the influence of longitudinal resolution variations, thus create a bias in the comparison of image quality for different reconstructed slice thicknesses. For this work, we designed a 'slice NEQ' that takes the longitudinal resolution into account and provides a volumetric indication of the scanner imaging performance. We also developed a specific formulation for the system DQE at the CT isocentre. A cylindrical water phantom with three different inserts for three contrast levels (−100, 120 and 1000 HU) was used for the task-based transfer functions (TTF) and noise power spectra (NPS) measurements. The longitudinal TTF was measured using the point source of the Catphan® 600 phantom. Images of the phantoms were acquired on two scanners (GE Discovery 750 HD and Siemens SOMATOM Force) and reconstructed using different slice thicknesses between 1 and 5 mm and algorithms (FBP for both systems, ASIR 50 and ASIR-V 50 for the GE). The slice NEQ correctly compared the imaging performance for different longitudinal resolutions whereas the 2D NEQ increased proportionally with the reconstructed slice thickness. The system DQE peaked at 0.70 (at 0.1 mm−1) for the Siemens and at 0.50 (at 0.1 mm−1) for the GE for FBP reconstructions. The validity of these Fourier-based metrics was restricted to a limited range of contrast due to nonlinearities introduced when dealing with iterative reconstructions (IR). (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1361-6560/ab807a

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
65
Journal Issue
10
Journal Page Range
[14 p.]
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
52072245
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ALGORITHMS; COMPUTERIZED TOMOGRAPHY; IMAGES; ITERATIVE METHODS; PERFORMANCE; PHANTOMS; POINT SOURCES; THICKNESS; TRANSFER FUNCTIONS
Descriptors DEC
CALCULATION METHODS; DIAGNOSTIC TECHNIQUES; DIMENSIONS; FUNCTIONS; MATHEMATICAL LOGIC; MOCKUP; RADIATION SOURCES; STRUCTURAL MODELS; TOMOGRAPHY